MP160 System
Highly Flexible System for Life Science Research
- Product No: NS-1065
- Manufacturer: Biopac
- EMG ECG EOG MRI Respiratory Airflow HR HRV
- Request Quote
Description
MP160 data acquisition and analysis systems with AcqKnowledge software provide a flexible tool for life science research. All systems are compliant with any Ethernet (UDP) ready computer running Windows or Mac. Record multiple data channels with variable sample rates to maximize storage efficiency at speeds up to 400 kHz (aggregate). Directly connect the computer to a single MP160 unit via the provided ETHUSB Ethernet adapter, or access multiple MP160 systems by connecting a switch box to the adapter*.
The base MP160 data acquisition system includes:
- Data acquisition unit: MP160
- Interface module: HLT100C
- AcqKnowledge® software CD
- License key (iLok USB) for AcqKnowledge 4.3+
- Software Guide (PDF)
- Ethernet Connection
- ETHUSB Ethernet adapter and Crossover Cable
- Power Supply
System Features
- Consolidated Hardware Support – MP160, MP150, MP36R, Mobita, BioHarness BT
- Linked acquisitions
- Focus Areas
- Data Integration – Noldus Observer X and SMI BeGaze
- Animal ECG Boundary Detection
Technical Specifications
Device
Max Sample Rate, MP Internal Memory: |
200 K samples/sec (400 K aggregate) |
PC Memory/Disk: |
200 K samples/sec (400 K aggregate) |
Internal Buffer: |
6 M samples |
Waveform Output Buffer: |
500 K samples |
Serial Interface Type/Rate: |
Ethernet: UDP (10M bits/sec) |
Transmission Type: |
Ethernet |
Maximum cable length: |
100 meters (Ethernet cable) |
Power Requirements: |
12 VDC @ 2 amp (uses AC150A) |
Dimensions: |
10 cm x 11 cm x 19 cm |
Weight: |
1.0 kg |
Operating Temperature Range: |
0-70° C |
Operating Humidity Range: |
0-95% |
OS Compatibility: |
Ethernet Interface: |
Analog Inputs
Number of Channels: |
16 |
Absolute Maximum Input: |
±15 V |
Operational Input Voltage: |
±10 V |
A/D Resolution: |
16 Bits |
Accuracy (% of FSR): |
±0.003 |
Input impedance: |
1.0 MΩ |
Analog Outputs
Number of Channels: |
2 |
Max output with acquisition: |
2 channels |
Output Voltage Range: |
±10 V |
D/A Resolution: |
16 bits |
Accuracy (% of FSR): |
±0.003 |
Output Drive Current: |
±5 mA (max) |
Output Impedance: |
100 |
Digital I/O
Number of Channels: |
16 |
Voltage Levels: |
TTL, CMOS |
Digital I/O Logic Type: |
CMOS |
Input Voltage Range: |
-0.5 V to 5.5 V (max) |
Input Clamp Current: |
±20 mA (max) |
Output Drive Current: |
±20 mA (max) |
External Trigger Input: |
TTL, CMOS compatible |
Logic Level Thresholds: |
Input Low Voltage: 1.50 V (max) |
Time Base
Min Sample Rate: |
2 samples/hour |
Trigger Options: |
Internal, External or Signal Level |
Power
Amplifier Module Isolation: |
Provided by the MP unit, isolated clean power |
CE Marking: |
EC Low Voltage and EMC Directives |
Leakage current: |
<8 µA (Normal), <400 µA (Single Fault) |
Fuse: |
2 A (fast below) |
Please note: Important! This product is for research applications only. Not a medical device as defined in EU directive 93/42/EEC. Not designed or intended to be used for diagnosis or treatment of disease.